Analysis of serum homocysteine in the laboratory practice - comparison of the direct chemiluminescence immunoassay and high performance liquid chromatography coupled with fluorescent detection

Introduction: Effective diagnosis of cardiovascular diseases requires the right tools to be used enabling selective and sensitive analysis of their biomarkers.One of them is homocysteine (Hcy), nowadays determined by immunoassays and chromatographic methods. This study aims to comparethe results obt...

Full description

Permalink: http://skupnikatalog.nsk.hr/Record/nsk.NSK01001102457/Details
Matična publikacija: Biochemia medica (Online)
30 (2020), 3 ; str. 439-445
Glavni autori: Paprotny, Łukasz (Author), Wianowska, Dorota, Izdebska, Magdalena, Celejewska, Agnieszka, Szewczak, Dorota, Solski, Janusz
Vrsta građe: e-članak
Jezik: eng
Predmet:
Online pristup: https://doi.org/10.11613/BM.2020.030703
Biochemia medica (Online)
Hrčak
LEADER 03531naa a22004214i 4500
001 NSK01001102457
003 HR-ZaNSK
005 20210719140812.0
006 m d
007 cr||||||||||||
008 210527s2020 ci d |o |0|| ||eng
024 7 |2 doi  |a 10.11613/BM.2020.030703 
035 |a (HR-ZaNSK)001102457 
040 |a HR-ZaNSK  |b hrv  |c HR-ZaNSK  |e ppiak 
041 0 |a eng  |b eng 
042 |a croatica 
044 |a ci  |c hr 
080 1 |a 61  |2 2011 
080 1 |a 577  |2 2011 
100 1 |a Paprotny, Łukasz  |4 aut  |9 HR-ZaNSK 
245 1 0 |a Analysis of serum homocysteine in the laboratory practice - comparison of the direct chemiluminescence immunoassay and high performance liquid chromatography coupled with fluorescent detection  |h [Elektronička građa] /  |c Łukasz Paprotny, Dorota Wianowska, Magdalena Izdebska, Agnieszka Celejewska, Dorota Szewczak, Janusz Solski. 
300 |b Graf. prikazi. 
504 |a Bibliografija: 27 jed. 
504 |a Summary. 
520 |a Introduction: Effective diagnosis of cardiovascular diseases requires the right tools to be used enabling selective and sensitive analysis of their biomarkers.One of them is homocysteine (Hcy), nowadays determined by immunoassays and chromatographic methods. This study aims to comparethe results obtained by direct chemiluminescence immunoassay (CLIA) and high performance liquid chromatography with fluorescent detection(HPLC-FD) using commercial kits.Materials and methods: Homocysteine concentration was determined in serum samples obtained from 101 individuals, using Atellica IM HCY(Siemens Healthineers, Erlangen, Germany) and HCY in plasma/serum – HPLC-FD (Chromsystems Instruments & Chemicals GmbH, Gräfelfing, Germany)tests validated for routine analysis. The latter was applied as a reference method. The comparability and agreement between the tested methodswere evaluated using the Passing-Bablok (PB) regression analysis and the Bland-Altman (BA) method of the differences analysis.Results: Studies showed that CLIA gives higher Hcy concentrations (15.7 ± 4.14 μmol/L). Passing-Bablok regression analysis of the results obtainedwith CLIA (y) compared with HPLC-FD (x) yielded an intercept of 0.22 (95%CI: - 2.16 to 2.46) and slope of 1.58 (95%CI: 1.33 to 1.87). Bland-Altmananalysis demonstrated a systematic positive bias for CLIA of 5.85 ± 2.77 μmol/L.Conclusions: Methods disagreement precludes their interchangeability. Lower Hcy values by HPLC-FD result from its greater selectivity. High performanceliquid chromatography with fluorescent detection should be considered as preferential method for analysing Hcy in blood serum as well asthe recommended reference method for routine clinical analysis. This fact, however, imposes the need to establish new reference ranges. 
653 0 |a Homocistein  |a Imunološki testovi  |a Luminiscentna mjerenja  |a Tekućinska kromatografija 
700 1 |a Wianowska, Dorota  |4 aut  |9 HR-ZaNSK 
700 1 |a Izdebska, Magdalena  |4 aut  |9 HR-ZaNSK 
700 1 |a Celejewska, Agnieszka  |4 aut  |9 HR-ZaNSK 
700 1 |a Szewczak, Dorota  |4 aut  |9 HR-ZaNSK 
700 1 |a Solski, Janusz  |4 aut  |9 HR-ZaNSK 
773 0 |t Biochemia medica (Online)  |x 1846-7482  |g 30 (2020), 3 ; str. 439-445  |w nsk.(HR-ZaNSK)000655372 
981 |b Be2020  |b B03/20 
998 |b dalo2107 
856 4 0 |u https://doi.org/10.11613/BM.2020.030703 
856 4 0 |u https://www.biochemia-medica.com/en/journal/30/3/10.11613/BM.2020.030703  |y Biochemia medica (Online) 
856 4 1 |y Digitalna.nsk.hr 
856 4 0 |u https://hrcak.srce.hr/244727  |y Hrčak